(1) The following minimal procedures shall be used in the fitting and selling of hearing aids:
(a) Pure tone audiometric testing by air and bone to determine the type and degree of hearing deficiency.
(b) Effective masking when indicated.
(c) Appropriate testing to determine speech reception thresholds, speech discrimination scores, the most comfortable listening levels, uncomfortable loudness levels, and the selection of the best fitting arrangement for maximum hearing aid benefit.
(2) The following equipment shall be used:
(a) A wide range audiometer which meets the specifications of the American National Standards Institute for diagnostic audiometers.
(b) A speech audiometer or a master hearing aid in order to determine the most comfortable listening level and speech discrimination.
(3) A final fitting ensuring physical and operational comfort of the hearing aid shall be made.
(4) The following medical clearance shall be obtained: If, upon inspection of the ear canal with an otoscope in the common procedure of a hearing aid fitter and upon interrogation of the client, there is any recent history of infection or any observable anomaly, the client shall be instructed to see a physician, and a hearing aid shall not be fitted until medical clearance is obtained for the condition noted. If, upon return, the condition noted is no longer observable and the client signs a medical waiver, a hearing aid may be fitted. Any person with a significant difference between bone conduction hearing and air conduction hearing must be informed of the possibility of medical correction.
(5)(a) A hearing aid office must have available, or have access to, a selection of hearing aid models, hearing aid supplies, and services complete enough to accommodate the various needs of the hearing aid wearers.
(b) At the time of the initial examination for fitting and sale of a hearing aid, the attending hearing aid specialist must notify the prospective purchaser or client of the benefits of telecoil, “t” coil, or “t” switch technology, including increased access to telephones and noninvasive access to assistive listening systems required under the Americans with Disabilities Act of 1990.
(6) Each audiometric test conducted by a licensee or authorized trainee in the fitting and selling of hearing aids shall be made in a testing room that has been certified by the department, or by an agent approved by the department, not to exceed the following sound pressure levels at the specified frequencies: 250Hz-40dB, 500Hz-40dB, 750Hz-40dB, 1000Hz-40dB, 1500Hz-42dB, 2000Hz-47dB, 3000Hz-52dB, 4000Hz-57dB, 6000Hz-62dB, and 8000Hz-67dB. An exception to this requirement shall be made in the case of a client who, after being provided written notice of the benefits and advantages of having the test conducted in a certified testing room, requests that the test be conducted in a place other than the licensee’s certified testing room. Such request shall be documented by a waiver which includes the written notice and is signed by the licensee and the client prior to the testing. The waiver shall be executed on a form provided by the department. The executed waiver shall be attached to the client’s copy of the contract, and a copy of the executed waiver shall be retained in the licensee’s file.
(7) The board shall have the power to prescribe the minimum procedures and equipment which shall be used in the conducting of hearing assessments, and for the fitting and selling of hearing aids, including equipment that will measure the hearing aid’s response curves to ensure that they meet the manufacturer’s specifications. These procedures and equipment may differ from those provided in this section in order to take full advantage of devices and equipment which may hereafter become available and which are demonstrated to be of greater efficiency and accuracy. The board shall adopt and enforce rules necessary to carry out the provisions of this subsection and subsection (6).
(8) Any duly authorized officer or employee of the department shall have the right to make such inspections and investigations as are necessary in order to determine the state of compliance with the provisions of this section and the applicable rules and may enter the premises of a licensee and inspect the records of same upon reasonable belief that a violation of this law is being or has been committed or that the licensee has failed or is failing to comply with the provisions of this act.
History.—ss. 11, 18, ch. 83-153; ss. 1, 7, ch. 84-94; s. 1, ch. 85-262; ss. 11, 19, 20, ch. 86-283; s. 22, ch. 90-341; s. 11, ch. 90-345; s. 4, ch. 91-429; ss. 250, 335, ch. 94-119; s. 8, ch. 94-160.
Structure Florida Statutes
Title XXXII - Regulation of Professions and Occupations
Chapter 484 - Dispensing of Optical Devices and Hearing AIDS
Part II - Fitting and Dispensing of Hearing AIDS (Ss. 484.0401-484.059)
484.042 - Board of Hearing Aid Specialists; membership, appointment, terms.
484.044 - Authority to make rules.
484.045 - Licensure by examination.
484.0501 - Minimal procedures and equipment.
484.051 - Itemization of prices; delivery of hearing aid; receipt, packaging, disclaimer, guarantee.
484.0513 - Cancellation by medical authorization; purchaser’s right to return.
484.053 - Prohibitions; penalties.
484.054 - Sale or distribution of hearing aids through mail; penalty.
484.056 - Disciplinary proceedings.
484.058 - Declaration of place of business; posting of license and notice.